Model Number 3701
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 11/27/2014
Event Type  Injury  
Manufacturer Narrative
(b)(4).This charger model was associated with a field correction.Manufacturer's evaluation: - corrective and preventive action (capa) investigation was performed.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 1 of 2; reference mfr report: 1627487-2015-01284.It was reported the patient complained he experienced pocket heating while charging his scs ipg.The patient stated he experienced a burning sensation on the skin and stopped using the scs system sometime after the heating started happening.A replacement charging system was sent to the patient to address the issue.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHARGING SYSTEM RECHARGEABLE IPG
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5016376
MDR Text Key23513229
Report Number1627487-2015-01283
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Other
Report Date 07/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2017
Device Model Number3701
Device Lot Number3579610
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/19/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/02/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age41 YR
-
-