TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 801763 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/30/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).After the case was over, they shut down the system, waited a couple seconds and restarted.The battery indicator normal (green) now after a full overnight charge.The customer has resolved the reported issue.Per follow-up with the ccp on 17-aug-2015: the system was in the hallway the day before it was pulled into room to use.Someone was there working on electronic records device.The ccp's guess was that he left the system on while unplugged.The ccp noticed the issue during her daily checklist and noticed it had already been on battery for unknown amount of time.When unplugged, noted that there was minimal/no battery backup available.The power light emitting diode (led) light on the front panel of the system-1 was red.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the customer found the system-1 completely dead (power switch in the on position).Then once plugged in for a while, battery indicator only showed ten minutes.The device was not changed out.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.The field service representative (fsr) did not need to go out to hospital, as the customer is aware of proper battery maintenance.No additional action will be taken at this time.
|
|