Model Number 23MJ-501
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problems Calcium Deposits/Calcification (1758); Rupture (2208)
Event Date 08/03/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Our investigation was limited to the review of the device history record, which showed that each manufacturing and inspection operation was performed and indicated complete in accordance with sjm specifications and procedures.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2015, a mitral valve replacement procedure was performed due to mitral stenosis in a patient with a heavily calcified annulus.The annulus was sized with a sjm 905 sizer set and measured between 21 mm and 23 mm.This 23 mm sjm masters series mechanical heart valve was selected and seated in the annulus without difficulty.Intraoperatively when coming off cardiopulmonary bypass, a rupture of the left ventricle (class1, blow-out type) occurred which required repair and the 23mm valve was removed.A smaller 21 mm sjm masters series mechanical heart valve was implanted.The procedure time was prolonged, but was completed successfully and the patient was reported to be stable postoperatively.According to the physician, the rupture of the left ventricle was caused by over-sizing as the 23 mm valve was too tight for the patient's annulus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES MECHANICAL HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key5018230
MDR Text Key23573295
Report Number2648612-2015-00017
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Report Date 08/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/10/2016
Device Model Number23MJ-501
Device Catalogue Number23MJ-501
Device Lot Number3439591
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/12/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-