MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE); IMPLANT, INTRAGASTRIC FOR MORBID OBESITY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE); IMPLANT, INTRAGASTRIC FOR MORBID OBESITY Back to Search Results
Catalog Number B-20265
Device Problems Calcified (1077); Inadequacy of Device Shape and/or Size (1583)
Patient Problems Nausea (1970); Pain (1994); Vomiting (2144); Patient Problem/Medical Problem (2688)
Event Date 11/11/2014
Event Type  Injury  
Manufacturer Narrative
Taper ii.The reporter of the complaint was asked to return the product for analysis.The device has not yet been received by allergan.Based upon the serial number and implant date provided by the reporter the connector type is assumed to be a taper ii.Visual examination may determine the connector type associated with this report.Allergan has not received the product at this time.Therefore no analysis or testing has been done.Band restriction issues, pain, vomiting, intolerance and nausea are surgical/physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.(b)(4).Device labeling addresses the reported event of band restriction issue as follows: "the lap-band ap system is available in two sizes, small and large.The physician should choose the appropriate size depending upon the patient's individual anatomy.After resolution of postoperative edema, most patients with appropriately placed bands report minimal if any restriction until saline is added to the band, regardless of the size used." device labeling addresses the potential for failure due to calcification as follows: ".The lap-band® system is not a lifetime product and it may break or fail, in whole or in part, at any time after implantation and notwithstanding the absence of any defect." device labeling addresses the reported events of pain, vomiting and nausea as follows: "band slippage and/or pouch dilatation can occur.Gastroesophageal reflux, nausea and/or vomiting with early or minor slippage may be successfully resolved by band deflation in some cases.More serious slippages may require band repositioning and/or removal.Immediate re-operation to remove the band is indicated if there is total stoma-outlet obstruction that does not respond to band deflation or if there is abdominal pain." device labeling addresses the contraindication for patients regarding intolerance as follows: "patients who are known to have, or suspected to have, an allergic reaction to materials contained in the system or who have exhibited pain intolerance to implanted devices.".
 
Event Description
Physician reported patient had intolerance to band and experienced nausea, vomiting and pain.Patient experienced band tightness and physician could not fill band.Explanted device showed calcifications on port band tubing.Device was not replaced and a roux-en-y bypass was done instead.Follow up findings: physician stated "patient very little compliant in treatment.Patient did not show up at several of the appointments given.".
 
Manufacturer Narrative
Supplement #1 - medwatch sent to fda on 02/25/2016.Device evaluation summary: the device was returned to apollo for analysis, and visual examination confirmed the connecter type as taper ii.A visual analysis of the returned device was performed and noted the returned device to be separated into three pieces, the access port, a piece of the band tubing joined by the ss connector, and the lap-band itself.The band tubing was noted to have openings.The tubing was also noted to have residue present on the outer surface.An air leak test was performed on the access port, and no leakage was noted.A microscopic analysis was then performed on the tubing, and noted two striated openings; the striations were consistent with a surgical end cut for removal of the device.Microscopic analysis noted one unidentified opening on the band tubing which was attached to the ss connector.The lap-band was air leak tested, and no bubbles or openings were noted.A fill test was then performed, and no blockage of the port was noted.Intolerance is a physiological complication and analysis of the device generally does not assist apollo in determining a probable cause for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND AP ADJUSTABLE GASTRIC BANDING SYSTEM (LARGE)
Type of Device
IMPLANT, INTRAGASTRIC FOR MORBID OBESITY
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia, costa rica, cs
CS  
Manufacturer Contact
laura leboeuf
1120 s. capital of texas hwy
bldg 1, ste 300
austin, TX 78746
5122795141
MDR Report Key5020919
MDR Text Key23688291
Report Number3006722112-2015-00261
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Followup
Report Date 07/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/26/2008
Device Catalogue NumberB-20265
Device Lot Number1332371
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/21/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/25/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age43 YR
-
-