Investigation conclusion update: it is indicated that product is not returning for evaluation.Since the product associated with the complaint was not returned, a review of in-house testing was performed.Retain strip testing results met both accuracy and repeatability criteria.The release specification is within the calculated confidence interval of the percent flier rate for complaint investigation testing.No product deficiency was found for lot k363299.Although a relevant nc was noted in batch record, it did not affect the final release specifications.There is no indication of a product deficiency and additional corrective actions were not required.Root cause could not be determined from the information provided by the customer.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.
|