ABON BIOPHARM (HANGZHOU) CO., LTD CONSULT DIAGNOSTICS HCG DIPSTICK; HCG PREGNANCY TEST Back to Search Results
Model Number FHC-101-OBC517
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Ectopic Pregnancy (1819)
Event Date 06/24/2015
Event Type  Injury  
Manufacturer Narrative
The customer's results were not replicated in-house with retention devices.Retention devices were tested with 25 miu/ml cutoff hcg urine control and 3 high level of hcg urine controls (205.2 iu/ml, 208.6 iu/ml, and 216.8 iu/ml); all results were positive at read time.No false negatives were obtained.Manufacturing batch record review did not uncover any deviations or non-conformances.The root cause of the customer's false negative results could not be determined from the provided information and without patient specimen in-house analysis.Based on the information available, there is no indication of a product deficiency.No corrective action is required at this time.This issue will be subject to tracking and trending.
 
Event Description
Caller reported a potential false negative urine hcg result with consult diagnostics hcg dipstick.After receiving the negative test result in the office, they placed an iud device in the patient for birth control.The patient left the office and a few weeks later she ended up in the hospital.The patient had a tubal pregnancy and needed surgery.She was 8 weeks pregnant at the time of the ectopic pregnancy surgery.The iud device was placed in the patient about 6 weeks before that, so she was about 2 weeks pregnant at the time of the iud placement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULT DIAGNOSTICS HCG DIPSTICK
Type of Device
HCG PREGNANCY TEST
Manufacturer (Section D)
ABON BIOPHARM (HANGZHOU) CO., LTD
#198 12th street east
hangzhou econ & tech dev area
hangzhou, zhejiang zhejiang 31001 8
CH  310018
Manufacturer (Section G)
ABON BIOPHARM (HANGZHOU) CO., LTD
#198 12th street east
hangzhou econ & tech dev area
hangzhou, zhejiang zhejiang 31001 8
CH   310018
Manufacturer Contact
ya-ling king
9975 summers ridge rd.
san diego, CA 92121
8588052084
MDR Report Key5022785
MDR Text Key23811680
Report Number2027969-2015-00613
Device Sequence Number1
Product Code JHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K993203
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other
Report Date 07/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2016
Device Model NumberFHC-101-OBC517
Device Lot NumberHCG4070214
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
-
-