• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; LARGE VOL. NEBULIZER HEATER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON THERMAGARD NEBULIZER HEATER; LARGE VOL. NEBULIZER HEATER Back to Search Results
Catalog Number 5705
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/12/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device sample was not returned for evaluation at the time of this report.Serial# reported as - (b)(4).
 
Event Description
The customer alleges that the unit is overheating during use.No patient injury or harm reported.
 
Manufacturer Narrative
(b)(4).According to the records of the vendor ((b)(4)), the unit in reference did not present any problems or malfunction during the production process.The heater was tested in accordance with (b)(4) procedure and was found to meet specifications.The estimated age of the unit is approximately 2278 days.The unit was not returned for evaluation; therefore, the complaint could not be confirmed.If the unit is returned, a follow-up report will be submitted with investigation results.
 
Event Description
The customer alleges that the unit is overheating during use.No patient injury or harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON THERMAGARD NEBULIZER HEATER
Type of Device
LARGE VOL. NEBULIZER HEATER
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
2917 weck drive
research triangle park NC 27709
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key5024405
MDR Text Key23827938
Report Number1044475-2015-00330
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Report Date 08/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number5705
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-