Model Number 302-20 |
Device Problems
Break (1069); Battery Problem (2885); Moisture or Humidity Problem (2986)
|
Patient Problem
No Code Available (3191)
|
Event Date 08/10/2015 |
Event Type
malfunction
|
Event Description
|
It was reported on (b)(6) 2015 that the patient was having a replacement for battery depletion.The impedance value was measured prior to the case and was ok.When the surgeon opened up the pocket he noticed that the lead was nicked and there was fluid inside of it.The devices were discarded after surgery and will not be returned for analysis.
|
|
Manufacturer Narrative
|
Describe event or problem; inadvertently reported that the devices were discarded after surgery.Evaluation codes, conclusion; inadvertently reported code that device was discarded.
|
|
Event Description
|
The generator and lead were not discarded and were returned for analysis on 08/25/2015.Analysis is currently underway but has not been completed to date.
|
|
Event Description
|
Product analysis for the generator was completed and approved on 09/23/2015.During analysis it was found that the generator was pulse disabled.Burn marks were observed on the pulse generator case, which indicated that the pulse generator may have been exposed to an electro-cautery tool during device explant.Review of the device decoder data downloaded from the generator shows that on (b)(6) 2015 (date of explant) the impedance value changed from 1764 ohms to 20879 ohms.The vns programming history database shows the last known diagnostic test was performed on (b)(6) 2015 with an impedance value of 1778 ohms.Other than the noted event (pulsedisabled), there were no additional performance or any other type of adverse conditions found with the pulse generator.Product analysis for the lead was completed and approved on 09/24/2015.The electrodes were not returned for analysis and therefore a complete evaluation could not be performed on the entire lead product.The abraded opening, incision and slice marks found on the outer silicone tubing, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.The condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No obvious anomalies were noted.Continuity checks of the returned lead portion were performed, during the visual analysis, and no discontinuities were identified.Based on the findings in lab, there is no evidence to suggest discontinuities in the returned portion of the device.
|
|
Event Description
|
Information was received on 05/13/2016 from operative notes from the date of explant (b)(6) 2015.It was stated that the lead to the vagal nerve stimulator had loss of integrity.There was fluid inside of the costing around the lead up and down the lead of the device.It was noted on inspection of the lead that there was fluid in the lead extending to the generator.They them identified an opening in the elastic tubing surrounding the lead wires which allowed for an ingress in fluid.With this finding, a decision was made to replace the lead.
|
|
Search Alerts/Recalls
|