Model Number EX1150
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that there was a burning smell from power plug.
 
Manufacturer Narrative
Investigation of the unit verified the alleged complaint by reproducing this specific issue.Root cause was broken pcb connector.The reason for this breakage happened when changing power supply prong.This interrupted the power flowing into the transformer and consequently interrupting the power into the unit.Since there is no physical connection anymore when trying to plug in the power supply back into an outlet all the voltage was being applied to the small broken connector and there was no way for the current to flow causing the broken piece to heat up and consequently start smelling like burnt.There is no patient safety risk as component is part of the power supply pcb board and it is highly protected by the power supply housing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (CRM-KISTA)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
ulla strindlund
box 7051
isafjordsgatan 15
kista SE-16-407
SW   SE-16407
4684744043
MDR Report Key5027168
MDR Text Key24354951
Report Number3010215456-2015-28260
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Followup
Report Date 07/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberEX1150
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/25/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received12/21/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-