It was reported that the vns patient underwent generator replacement surgery on (b)(6)2015 due to desire for the latest generator model.During the procedure, the surgeon observed fluid in the lead insulation and twisting of the lead at the generator pocket possibly due to patient manipulation.Diagnostic tests with the replacement generator and existing lead showed lead impedance within normal limits (impedance value - 3205 ohms).The surgeon elected not to replace the lead and untangled the lead before completing surgery.The explanted generator has been returned to the manufacturer where analysis is currently underway.
|