Patient had pulmonary edema with a considerable amount of fluid in the tubing.Rn returned to find pt.Being suctioned and oscillator alarming and not running.The differential pressure (delta p) was reading 0 cmh2o and the mean airway pressure (map) was 112 cmh2o.The oscillator would not restart after several attempts.Biomed was able to verify the device would not start.After troubleshooting further, it was noticed that the patient sensing line filter was creating the issue.With the filter removed from the line, the oscillator restarted and functioned normally.The pt.Was a (b)(6) on the adult oscillator (3100b) with lower than usual settings, but still within normal operating range.Typical adult delta p is 80 and this pt.Had 60.It was determined that either the sensing line pressure or a saturated filter was the root cause for stopping the oscillator.Biomed performed tests on the device to ensure it was operating correctly.Respiratory updated their process to swap out filter every 24 hours.This will mitigate a wet filter stopping the oscillator or moisture reaching and damaging the transducer circuit board.Manufacturer response for high frequency oscillating ventilator, 3100b high-frequency oscillatory ventilator (per site reporter): as of this report, the manufacturer has not responded.
|