(b)(4).Manufacturing site evaluation: samples received: 1 bm067r mayo-hegar needle holder.An investigation was performed visually.The metal inlay of one jaw was broken off and was not returned for evaluation.The broken segment is approximately 2 mm from the tip of the jaw.The labeling of the instrument was not completed by aesculap.Also the solder joint is not to aesculap standards.The device appears to have been repaired by a foreign company and does not meet aesculap standards.Final conclusion: complaint is not justified.The failure result is user related, using non authorized repair service.Corrective/preventive actions: not applicable.
|