• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS PROGRAMMING SOFTWARE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS PROGRAMMING SOFTWARE Back to Search Results
Model Number MODEL 250
Device Problem Use of Incorrect Control/Treatment Settings (1126)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/21/2015
Event Type  malfunction  
Event Description
During review of the patient's clinic notes , it was identified that a programming anomaly occurred.It is suspected that the system diagnostic test performed on (b)(6) 2015 had faulted and changed the patient's vns settings to unintended settings.It is unknown if final interrogation was performed on (b)(6) 2015.Manufacturer labeling indicates to perform a final interrogation to confirm intended settings prior to patients leaving the clinic.Upon initial interrogation on the subsequent visit, faulted settings were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
CYBERONICS
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key5033733
MDR Text Key25035669
Report Number1644487-2015-05624
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 08/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 250
Device Lot Number1077277
Other Device ID NumberVERSION 8.1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/26/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/08/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age61 YR
-
-