Model Number 10-85-00 |
Device Problems
Device Inoperable (1663); Other (for use when an appropriate device code cannot be identified) (2203)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/27/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Sorin group (b)(4) manufactures the s5double head pump.The incident occured in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that post-operatively, the pump was not functioning.There was no patient involvement.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group rec'd a report that post-operatively, the pump was not functioning.There was no patient involvement.
|
|
Manufacturer Narrative
|
Livanova (b)(4) manufactures the s5 double head pump.The incident occurred in (b)(6).(b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to confirm the reported issue and traced the failure to a defective shaft angle encoder.The shaft angle encoder was replaced to resolve the reported issue.The defective component was returned to livanova (b)(4) for further investigation and a defective shaft encoder bearing was defective.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
|