Model Number SC-2218-50 |
Device Problems
Bent (1059); Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 07/20/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model #: sc-2218-50, serial #: (b)(4), description: linear st lead, 50cm.
|
|
Event Description
|
A report was received that the patient was not getting stimulation due to several contacts out on each lead.Database analysis confirmed the impedance measurements revealed high values on several contacts.The patient will undergo leads replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that the patient underwent a replacement procedure and did well postoperatively.Sc-2218-50 sn (b)(4): device evaluation indicated that leads complaint had been confirmed.Visual (microscope) and x-ray inspection of the lead revealed that 5 cables were fractured at the bent/kinked location of the lead.The bent/kinked location is 1 cm from the set screw mark of the clik anchor.The fractured cables resulted in the reported high impedance complaint.No cables were exposed at the fracture site.Sc-2218-50 sn (b)(4): device evaluation indicated that the lead had no anomalies or deviations were found during the complaint investigation site (cis) review, there was no reason to suspect a manufacturing defect as the source of the reported complaint.
|
|
Event Description
|
A report was received that the patient was not getting stimulation due to several contacts out on each lead.Database analysis confirmed the impedance measurements revealed high values on several contacts.The patient will undergo leads replacement procedure.
|
|
Search Alerts/Recalls
|