SORIN GROUP DEUTSCHLAND S5 DOUBLE HEAD PUMP ; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-85-00
Device Problems Device Inoperable (1663); Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/30/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 double head pump.The incident occurred in cedex, france.This medwatch report is filed on behalf of sorin group (b)(4).The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group received a report that the touch screen of the s5 double head pump was unresponsive during setup.There was no report of patient injury.
 
Manufacturer Narrative
There have been two field actions for this issue.The z-numbers are z-0956/0965-2011 and z-1149/1158-2012.(b)(4) manufactures the s5 double head pump.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was unable to reproduce the reported issue and did not identify any signs of liquid ingress into the pump panel or the touch screen.The touch screen, hkr board and ribbon cables were replaced to resolve the issue.The pump software was then updated to rev 16 per the user-specified parameters.Subsequent functional testing did not identify further issues and the unit was returned to service.The date code of the kapton-tail connection of the replaced touch screen suggests that the root cause is aging and wear of the device and glue.This is a known issue and a root cause investigation (b)(4) has already been investigated.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 DOUBLE HEAD PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munchen D 809 39
GM  D 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5043092
MDR Text Key24688066
Report Number9611109-2015-00312
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Followup
Report Date 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-85-00
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/27/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0956/0965-2011
Patient Sequence Number1
-
-