The incident device anticipated to return.A follow-up report will be provided upon completion of investigation.In accordance to 21 cfr 803.56, if we obtain additional information, which was not known or was not available when the initial report was submitted, then the supplemental report will be submitted to the fda.
|
(b)(4).Investigation summary: the event unit was returned for evaluation.Upon inspection, engineering confirmed the dent on the shaft.The dent on the shaft added friction to the shaft; which likely resulted in the rough actuation and short feeding as observed by the customer.The root cause of the damaged shaft likely happened while removing the device from the tray packaging.All clip appliers undergo 100% visual and functional inspection during the manufacturing and assembly process.As a part of this process, applied medical is currently investigating packaging enhancements intended to further minimize the potential for this type of incident to occur.In accordance to 21 cfr 803.56, if we obtain additional information, which was not known or was not available when this report was submitted, then the supplemental report will be submitted to the fda.
|