Model Number CA500
Device Problems Bent (1059); Mechanics Altered (2984)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/06/2015
Event Type  malfunction  
Manufacturer Narrative
The incident device anticipated to return.A follow-up report will be provided upon completion of investigation.In accordance to 21 cfr 803.56, if we obtain additional information, which was not known or was not available when the initial report was submitted, then the supplemental report will be submitted to the fda.
 
Event Description
Laparoscopic hernia- "the circulator nurse noticed that the ca500 had a dent in it but opened it anyway.When they attempted to load the clip it came out sideways in the jaw of the device.They opened another ca500 and finished the case." type of intervention: "na." patient status: "fine.".
 
Manufacturer Narrative
(b)(4).Investigation summary: the event unit was returned for evaluation.Upon inspection, engineering confirmed the dent on the shaft.The dent on the shaft added friction to the shaft; which likely resulted in the rough actuation and short feeding as observed by the customer.The root cause of the damaged shaft likely happened while removing the device from the tray packaging.All clip appliers undergo 100% visual and functional inspection during the manufacturing and assembly process.As a part of this process, applied medical is currently investigating packaging enhancements intended to further minimize the potential for this type of incident to occur.In accordance to 21 cfr 803.56, if we obtain additional information, which was not known or was not available when this report was submitted, then the supplemental report will be submitted to the fda.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CA500, EPIX UNIVERSAL CLIP APPLIER 3/BX
Type of Device
FZP
Manufacturer (Section D)
APPLIED MEDICAL
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer (Section G)
APPLIED MEDICAL
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
22872 avenida empresa
rancho santa margarita, CA 92688
9497138233
MDR Report Key5043586
MDR Text Key25284460
Report Number2027111-2015-00613
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011236
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Followup
Report Date 08/31/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/03/2018
Device Model NumberCA500
Device Catalogue Number101281901
Device Lot Number1247833
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Hospital
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/31/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received01/18/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-