• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-017
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/04/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of this investigation confirmed the amplatzer septal occluder met all functional and dimensional specifications when analyzed at sjm.A review of the device history record confirmed the occluder met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the reported event remains unknown.
 
Event Description
After a 17 mm amplatzer septal occluder (aso) was deployed, it was determined to be the wrong size and was removed percutaneously with a snare.A 26 mm aso was implanted and the patient was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key5046573
MDR Text Key24734744
Report Number2135147-2015-00096
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 08/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Model Number9-ASD-017
Device Catalogue Number9-ASD-017
Device Lot Number4082408
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/01/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age17 YR
Patient Weight51
-
-