Model Number 109681-001 |
Device Problem
Crack (1135)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/05/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that during unpacking the catheter was cracked.This angiojet solent catheter was selected; however, during unpacking it was noticed that the catheter was cracked midway on the catheter.It was noted that the crack was at a portion that would enter the patient.The device never entered the patient.The procedure was completed with another of the same device.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: a visual and tactile examination of the returned complaint device revealed that the outer shaft had a kink in the tubing approximately 37cm from the tip of the catheter.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
|
|
Event Description
|
It was reported that during unpacking the catheter was cracked.This angiojet solent catheter was selected; however, during unpacking it was noticed that the catheter was cracked midway on the catheter.It was noted that the crack was at a portion that would enter the patient.The device never entered the patient.The procedure was completed with another of the same device.
|
|
Search Alerts/Recalls
|