• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS HEALTHCARE ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS HEALTHCARE ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number 722003
Device Problems Loose or Intermittent Connection (1371); Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
When investigation is completed a follow up report will be sent to fda.(b)(4).
 
Event Description
Philips received a complaint from a customer in which they explained the following problem: the customer moved the c-arm by pushing the control, when she lifted her finger off the control panel, the c-arm momentarily kept moving.The doctor felt that the safety glass was pushed by the c-arm too close to him and backed away but the sheath came loose as he did this.He then had to pull it out and press on the groin to stop the bleeding.No patient or user harm was reported due this issue.
 
Manufacturer Narrative
A philips field service engineer found that the table side operation joystick was sticking and therefore returning slowly to its zero position.This caused the movement to continue after it had been released by the customer.The fse cleaned the geo module mechanically, after which no further faults occurred.We do not consider this to be a structural issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS HEALTHCARE
veenpluis 4-6
p.o. box 10.000
best 5680 DA
NL  5680 DA
Manufacturer Contact
dusty leppert
veenpluis 4-6
p.o. box 10.000
best 5680 -DA
NL   5680 DA
MDR Report Key5047215
MDR Text Key25357649
Report Number3003768277-2015-00073
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Followup
Report Date 08/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number722003
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/01/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/03/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-