Model Number 840 |
Device Problems
Sticking (1597); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The field service engineer (fse) evaluated the ventilator and verified the reported issue.The fse replaced the pressure solenoid (psol) valve, which resolved the reported issue.The ventilator then passed all testing.
|
|
Event Description
|
It was reported that a ventilator displayed a message that the pressure solenoid (psol) valve was stuck.There was no report of patient involvement.There is no report of serious injury or death associated with this event.
|
|
Manufacturer Narrative
|
An investigation was performed and the technician found that the reported problem could not be duplicated.No failure was detected.If information is provided in the future, a supplemental report will be issued.
|
|