COVIDIEN 840 VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number 840
Device Problems Sticking (1597); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The field service engineer (fse) evaluated the ventilator and verified the reported issue.The fse replaced the pressure solenoid (psol) valve, which resolved the reported issue.The ventilator then passed all testing.
 
Event Description
It was reported that a ventilator displayed a message that the pressure solenoid (psol) valve was stuck.There was no report of patient involvement.There is no report of serious injury or death associated with this event.
 
Manufacturer Narrative
An investigation was performed and the technician found that the reported problem could not be duplicated.No failure was detected.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
thom mcnamara
15 hampshire street
mansfield, MA 02048
5084524811
MDR Report Key5047318
MDR Text Key25340982
Report Number8020893-2015-00834
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Respiratory Therapist
Type of Report Followup
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number840
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/28/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/01/2015
Supplement Dates Manufacturer Received08/04/2015
Supplement Dates FDA Received12/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/09/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-