It was reported that during use, the catheter clamp and the blue clamp were slightly detached from the catheter.As a result, the catheter was slightly pulled out of the patient.The md saw the patient touching it over the dressing, however, did not suspect it was the cause of the complaint.A new kit was opened and used without issue.There was no reported delay, death, or complications to the patient as a result of this occurrence.
|
(b)(4).Device evaluation: the reported complaint could not be confirmed.The customer returned one catheter along with a box clamp and fastener.The sample showed signs of use and the distal tip of the catheter was cut off and not returned.Further examination revealed that the box clamp and fastener were present on the catheter body, but no sutures were present on the juncture hub or clamp upon return.The od of the catheter met specification.The id of the clamp could not be measured since the material is pliable and it had been used.However, the clamp and fastener were reassembled on the catheter body and were found to be secure.A review of manufacturing records did not yield any relevant findings.The provided instructions state that the juncture hub is the primary suture site and the clamp is the secondary.It was not reported if the hub was sutured and no evidence of use was found on the suture wings.No problem was found with the returned sample.No further action will be taken.
|