TERUMO CARDIOVASCULAR SYSTEMS CORP. CDI BLOOD PARAMETER MONITORING SYSTEM 500; MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS-CDI BPM SYSTEM 500
|
Back to Search Results |
|
Model Number 500AVHCT |
Device Problem
Electrical Overstress (2924)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported complaint was confirmed.The aux pcba will be replaced in service.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The service repair technician (srt) reported that during laboratory analysis of the device at the product surveillance laboratory, the aux printed circuit board assembly (pcba) has capacitor c112 that has shorted or opened up and was destroyed.There was no patient involvement.
|
|