MAQUET CARDIOPULMONARY AG HEART LUNG MACHINE- HL20; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number MCP00703309
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
September 04, 2015 09:51 am (gmt-4:00) added by (b)(6): (b)(4).A maquet service technician evaluated the device and could not reproduce the reported issue.The device is being returned to the repair depot and a loaner device is being provided.A supplemental medwatch will be submitted when additional information becomes available.
 
Event Description
It was reported that the customer received the message rpm-diff and the pump would not work on some occasions.No reported patient effect.(b)(4).
 
Manufacturer Narrative
(b)(4).Field safety technician has been sent for investigation and could not duplicate issue - operated unit for one week with numerous warm and cold starts, checked calibration including belt tension and tacho phase,no issues found.Calibration and functional test as per the service manual.It was decided to replace the motor control board as a precaution.Failure could not be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE- HL20
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
MICHAEL CAMPBELL
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
MDR Report Key5057001
MDR Text Key25738323
Report Number8010762-2015-01035
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMCP00703309
Device Catalogue Number701027652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/05/2015
Initial Date FDA Received09/04/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-