Model Number MCP00703309 |
Device Problems
Device Inoperable (1663); Device Displays Incorrect Message (2591)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
September 04, 2015 09:51 am (gmt-4:00) added by (b)(6): (b)(4).A maquet service technician evaluated the device and could not reproduce the reported issue.The device is being returned to the repair depot and a loaner device is being provided.A supplemental medwatch will be submitted when additional information becomes available.
|
|
Event Description
|
It was reported that the customer received the message rpm-diff and the pump would not work on some occasions.No reported patient effect.(b)(4).
|
|
Manufacturer Narrative
|
(b)(4).Field safety technician has been sent for investigation and could not duplicate issue - operated unit for one week with numerous warm and cold starts, checked calibration including belt tension and tacho phase,no issues found.Calibration and functional test as per the service manual.It was decided to replace the motor control board as a precaution.Failure could not be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
|
|
Event Description
|
(b)(4).
|
|