Model Number 3788
Device Problem Temperature Problem (3022)
Patient Problems Burning Sensation (2146); Discomfort (2330)
Event Date 08/07/2015
Event Type  Injury  
Manufacturer Narrative
This ipg serial number was included in a field correction.(b)(4).
 
Event Description
Device 1 of 2.Reference mfr.Report#: 1627487-2015-23512.It was reported the patient experienced uncomfortable heating at the ipg site while recharging.The patient was sent a replacement le charging system to address the issue.Follow-up information revealed the patient received the new charging system which resolved the reported issue.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5058219
MDR Text Key25097725
Report Number1627487-2015-23511
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Other
Report Date 08/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2012
Device Model Number3788
Device Lot Number3088846
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/04/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age53 YR
-
-