MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI CHARGING SYSTEM; SCS CHARGER Back to Search Results
Model Number 3721
Device Problem Temperature Problem (3022)
Patient Problems Burning Sensation (2146); Discomfort (2330)
Event Date 08/06/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).This charger model was associated with a field correction.Manufacturer's evaluation: corrective and preventive action (capa) investigation was performed.The investigation for capa (b)(4) associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.
 
Event Description
Device 1 of 2.Reference mfr report: 1627487-2015-01314.It was reported the patient experiences heating while charging his scs ipg.A new charging system was sent to the patient to address the issue.On 08/01/2012 st.Jude medical, neuromodulation division, sent field action letters to patients related to heating while charging and raised awareness of this issue to patients.An increase in prior non-reported heating while charging events and other non-reported events was expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI CHARGING SYSTEM
Type of Device
SCS CHARGER
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5058251
MDR Text Key25097605
Report Number1627487-2015-01313
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Other
Report Date 08/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date08/28/2011
Device Model Number3721
Device Lot Number3178855
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/04/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/09/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age45 YR
-
-