Model Number 10-80-00 |
Device Problems
Blank screen (1063); Device Inoperable (1663)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 08/05/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the roller pump touch screen was unresponsive during set up.There was no pt involvement.The investigation is ongoing.A f/u report will be sent when the investigation is complete.
|
|
Event Description
|
Sorin group received a report that the roller pump touch screen was unresponsive during set up.There was no pt involvement.
|
|
Manufacturer Narrative
|
(b)(4) manufactures the s5 roller pump.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to reproduce the reported issue.The touch screen and accessories were replaced and the nvmem was cleared.Subsequent functional testing found no further issues and the unit was returned to service.No further occurrences have been reported for this unit.Photos of the kapton-tail connection of the touch screen were provided to livanova (b)(4) for evaluation.The photos showed signs of fluid ingress into the touch screen, which led to oxidation of contacts and electrical failure.This is a known issue for which a capa (b)(4) has already been implemented.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
|
|