Model Number 10-80-00
Device Problems Blank screen (1063); Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 08/05/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the roller pump touch screen was unresponsive during set up.There was no pt involvement.The investigation is ongoing.A f/u report will be sent when the investigation is complete.
 
Event Description
Sorin group received a report that the roller pump touch screen was unresponsive during set up.There was no pt involvement.
 
Manufacturer Narrative
(b)(4) manufactures the s5 roller pump.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).A livanova field service representative was dispatched to the facility to investigate.The service representative was able to reproduce the reported issue.The touch screen and accessories were replaced and the nvmem was cleared.Subsequent functional testing found no further issues and the unit was returned to service.No further occurrences have been reported for this unit.Photos of the kapton-tail connection of the touch screen were provided to livanova (b)(4) for evaluation.The photos showed signs of fluid ingress into the touch screen, which led to oxidation of contacts and electrical failure.This is a known issue for which a capa (b)(4) has already been implemented.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich D 809 39
GM  D 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5060686
MDR Text Key25224998
Report Number9611109-2015-00323
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Unknown
Remedial Action Other
Type of Report Followup
Report Date 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/01/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/20/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-