MEDTRONIC NEUROMODULATION INTERSTIM II; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
|
Back to Search Results |
|
Model Number 3058 |
Device Problems
Migration or Expulsion of Device (1395); Unable to Obtain Readings (1516)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant product: product id: neu_unknown_lead, implanted: (b)(6) 2015, product type: lead.(b)(4).
|
|
Event Description
|
It was reported that there was a replacement/revision.A lead issue with migration/dislodgement was reported.There was an invalid impedance issue.During the lead revision the representative reran impedance at 2 and 300pw (pulse width) and got c, 1 at ??? and c, 2 at ???.The representative reran at 2 volts and 360pw and all values came back within range.Caller needed to complete case.The revision had occurred.Case was in progress while call took place.There was no allegation of system use issue during the revision.Additional information has been requested but was not available as of the date of this report.A follow-up report will be made if additional information becomes available.
|
|
Search Alerts/Recalls
|
|
|