Lead management case to extract one non-functional bsc 0184 cardiac lead.The physician prepped the lead with an lld and began lasing with a 14f glidelight.Progress stalled and the device was upsized to a 16f glidelight.While the laser sheath was in the ra the lead release from the rv.A drop in bp was noted and upon visualization on tee a tear was identified on the tricuspid valve leaflet.The patient was stabilized with medication and transported to icu for observation.No additional treatments were required and the patient was discharged.Upon visual inspection of the icd lead, a significant amount of tissue was found attached to the lead along with extruded cables.The physician believes the primary cause of the tear was due to lead insulation and conductor cables pulling free from the leaflet during removal.The glidelights involved in this case were not in use near the tricuspid valve.This report is to reflect on the lld as it was the traction platform in use during lead removal.
|