The system was used for treatment.A review of kit lot d310 was performed.There were no nonconformances related to this complaint.This lot met release requirements.The uvadex lot number was not provided, since uvadex was not administered.However, a review of all uvadex lots manufactured since (b)(4) 2013 was performed.No trends or nonconformances related to the complaint were noted.Trends were reviewed for complaint categories, alarm #18: system pressure, air leak, and pressure dome membrane leak.A downward trend was identified for alarm #18: system pressure.No trend was identified for pressure dome membrane leak.A corrective and preventive action was initiated for complaint category, pressure dome membrane leak.Analysis of the returned photos is still in progress at the time of this report.A supplemental report will be filed when this analysis is complete.(b)(4).
|
Photos were returned for analysis.Based on the review of the photos, a leak at the collect pressure dome assembly was confirmed, not at the membrane, but at the joint between the pressure dome and the clear tubing from the pto.The root cause of the blood leak is likely a manufacturing operator error when forming the bond between the pto and the collect pressure dome, by not fully inserting the tubing.As part of the corrective action a quality notice was posted and trained to at the manufacturer on 4/29/15 to reaffirm solvent bonding at this site.Complaints of this nature are monitored through tracking and trending.Should a trend arise, further action will be taken.Based on the product return investigation, the complaint category was changed from pressure dome membrane leak to tubing leak.Trends were reviewed for complaint category, tubing leak, and no trends were detected.No corrective and preventive action was initiated for this complaint category.(b)(4).
|