Model Number 3711
Device Problem Temperature Problem (3022)
Patient Problem Burning Sensation (2146)
Event Date 06/24/2015
Event Type  Injury  
Manufacturer Narrative
(b)(4).This charger model was associated with a field correction.Manufacturer's evaluation: corrective and preventive action (capa) investigation was performed.Evaluation codes: results: pocket heating was confirmed.The investigation for capa (b)(4)associated with heating while charging (pocket heating) concluded that the charger was capable of transferring energy to the ipg at a rate that would cause heating of the ipg and/or charging wand of sufficient elevated temperature to cause pain and burns.The heating while charging was determined to be exacerbated by off-axis charging of shallow implanted ipgs and that all chargers were capable of elevated heating.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 2 of 2.Reference mfr.Report #1627487-2015-06334.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOBILE CHARGING SYSTEM
Type of Device
SCS CHARGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key5074238
MDR Text Key25648606
Report Number1627487-2015-06335
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 08/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2012
Device Model Number3711
Device Lot Number3370721
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2015
Initial Date Manufacturer Received 08/24/2015
Initial Date FDA Received09/14/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/18/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-07262012-001-C
Patient Sequence Number1
Treatment
MODEL 3244, SCS LEAD
Patient Outcome(s) Other;
Patient Age65 YR
-
-