MAUDE Adverse Event Report: SORIN GROUP USA, INC. CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP USA, INC. CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Catalog Number CA-20070
Device Problems Partial Blockage (1065); Other (for use when an appropriate device code cannot be identified) (2203)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/19/2015
Event Type  Other  
Manufacturer Narrative
Pt info was not provided.Sorin group received a report that during cardioplegia delivery there was no fluid coming out the cardioplegia tubing.The clinical specialist noted that where the luer connects to the tubing, it appeared blocked.There was no report of pt injury.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group received a report that during cardioplegia delivery there was no fluid coming out the cardioplegia tubing.The clinical specialist noted that where the luer connects to the tubing, it appeared blocked.There was no report of pt injury.
 
Manufacturer Narrative
Sorin group received a report that during cardioplegia delivery there was no fluid coming out the cardioplegia tubing.The clinical specialist noted that where the luer connects to the tubing, it appeared blocked.There was no report of patient injury.The involved cardioplegia wye line was returned to sorin group usa for evaluation.A visual inspection of the line confirmed that the female luer was completely occluded with the adhesive used for bonding.No other defects or abnormalities were found.There is no remaining inventory of this lot number.The reported issue was caused by a manufacturing error.The individuals responsible for assembling and inspecting this line were gathered together, the manufacturing operating procedure was reviewed, and the possible causes for this occlusion were discussed to ensure that the proper steps needed for this build were clearly understood.Sorin group usa has determined that a capa is not needed.There was no patient injury, no trend has been identified and a correction has been performed.Sorin group will continue to monitor the market for trends related to this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOPLEGIA ADAPTERS, VESSEL CANULAE, AND AORTIC ROOT CANNULAE
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP USA, INC.
14401 west 65th way
arvada CO 80004
Manufacturer (Section G)
SORIN GROUP USA, INC.
14401 w 65th way
arvada CO 80004
Manufacturer Contact
carrie wood
14401 w 65th way
arvada, CO 80004
3034676461
MDR Report Key5075021
MDR Text Key26006337
Report Number1718850-2015-00356
Device Sequence Number1
Product Code DWF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Remedial Action Other
Type of Report Followup
Report Date 08/19/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Catalogue NumberCA-20070
Device Lot Number1507900104
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/11/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/20/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-