• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION CAREFUSION; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems No Audible Alarm (1019); Failure to Cycle (1142)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/07/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The carefusion field service engineer evaluated the ventilator and was unable to duplicated the reported failure.The unit was checked thoroughly for pressures, settings, alarms, etc and no issues were found.Unit passed all company specs using standard ridged circuit and also with the flexible circuit.No reason for the failure to alarm could be found.(b)(4).
 
Event Description
The customer reported that while in patient use the ventilator shut down and the unit did not alarm.Pt was placed on another 3100a, no patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREFUSION
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
CAREFUSION
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
jill rittorno
75 n. fairway drive
vernon hills, IL 
8473628056
MDR Report Key5078235
MDR Text Key25880805
Report Number2021710-2015-01646
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 08/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/17/2015
Initial Date FDA Received09/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-