Model Number M-4800-01 |
Device Problem
Image Display Error/Artifact (1304)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/27/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
The hardware investigation has begun but it has not been completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.(b)(4).
|
|
Event Description
|
It was reported that a patient underwent an atrial fibrillation procedure with a carto 3 system and a visitag issue occurred.It was reported that the visitags are disappearing as new ones are acquired during radiofrequency and as the map is turned.Part of the visitags were shown but the visitags on the ridge were not.The carto app was reloaded again and the issue did not resolve.The procedure was continued and there were no patient consequence.The visitag issue is indicative of a reportable event as the visitags disappearing may potentially lead to patient injury.
|
|
Manufacturer Narrative
|
(b)(4).It was reported that a patient underwent an atrial fibrillation procedure with a carto 3 system and a visitag issue occurred.Bwi representative was informed that global preset (setting) of visitags is used for the entire study.If it is changed, all previous sessions are updated accordingly and it's not unique to each map.The issue was related to user error.The system is ready for use.The history of customer complaints associated with this carto 3 system was reviewed.There was not any additional reported complaints that may be related to the reported issue.Device history record review was performed by the manufacturer and no anomalies were noted in manufacturing or servicing of this equipment.
|
|