The subject device was returned to olympus.Olympus confirmed that the reported phenomenon was reproduced and the device worked properly when replacing the internal circuit board of the device to a spare circuit board.Olympus will perform further investigation to determine the cause of this phenomenon.Olympus also checked the manufacturing history of the subject device, and there was no irregularity found.The issue is under investigation, so olympus will submit a supplemental mdr report after the cause of this phenomenon is determined.This report is being submitted as a medical device report in an abundance of caution.
|
Omsc tried to reproduce the subject phenomenon, and the phenomenon was reproduced.Omsc found that the phenomenon was eliminated after replacing the internal circuit board of the subject uhi-4 to a spare circuit board.Omsc investigated this circuit board, but could not identify the failure part.Omsc evaluated the subject uhi-4 and found the following: there was not any abnormality about the behavior of the pressure sensor; there was not any abnormality about the behavior of the cavity pressure control.Based on the above investigation, omsc surmised that the subject phenomenon caused by the following: any adventitious failure occurred in the devices on the circuit board for controlling the insufflation rate.As a result, the subject uhi-4 insufflated air less than the setting value.The uhi-4 instruction manual states the corresponding method when there is an abnormality.There were no further details provided.If significant additional information is received, this report will be supplemented.
|