Model Number 48-30-00
Device Problem Device Inoperable (1663)
Patient Problem No Patient Involvement (2645)
Event Date 07/28/2015
Event Type  malfunction  
Manufacturer Narrative
Sorin group (b)(4) manufactures the s5 system.The incident occurred in (b)(6).This report is filed on behalf of sorin group (b)(4).Sorin group received a report that during set-up the pump touch screen was unresponsive.There was no patient involvement.The investigation is on-going.A follow-up report will be submitted when the investigation is complete.
 
Event Description
Sorin group received a report that during set-up the pump touch screen was unresponsive.There was no patient involvement.
 
Manufacturer Narrative
There have been two field actions for this issue.The z-numbers are z-0956/0965-2011 and z-1149/1158-2012.(b)(4) manufactures the s5 system.The incident occurred in (b)(6).This medwatch report is being filed on behalf of (b)(4).A (b)(4) field service representative was dispatched to the facility to investigate.The service representative tested the unit and was able to reproduce the error.The faulty touch screen was replaced with a new led touch screen.The unit was functionally tested without issue and returned to service.Photos of the touch screen were taken and provided to (b)(4) for further analysis.It was confirmed that the touch screen falls under an existing root cause investigation.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.This is a known issue for which a root cause investigation ((b)(4)) was initiated.Evaluated on site by (b)(4) technician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN S5 SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindbergh strasse 25
munchen D 809 39
GM  D 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5088138
MDR Text Key26259678
Report Number9611109-2015-00361
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Remedial Action Replace
Type of Report Followup
Report Date 02/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number48-30-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/15/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0956/0965-2011
Patient Sequence Number1
-
-