The device was received by the manufacturer with the nose cone severely damaged.It could not be determined if all of the material was present due to the distortion and melting of the sheath.The microtip was side-loaded during operation and the friction resulted in heat build up and subsequent damage to the outer nose cone.It is highly probable, based upon the fracturing of the device, that the microtip made contact with hard tissue.The damage to the device is attributed to improper technique by the user.The device was not functionally tested as the handpiece had been separated from the tubing and cartridge.Only the handpiece was received.Visual evaluation was conducted on the returned portion of the device.
|