SORIN GROUP DEUTSCHLAND STOCKERT CENTRIFUGAL PUMP CONSOLE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 60-03-00
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/15/2015
Event Type  malfunction  
Manufacturer Narrative
There was no patient information provided.Sorin group (b)(4) manufactures the stockert centrifugal pump console.The incident occurred in (b)(6).This medwatch report is filed on behalf of sorin group (b)(4).Sorin group received a report that the touch screen of an scpc centrifugal pump console was defective.There was no report of patient injury.A sorin group service representative was dispatched to the facility to evaluate the issue.The service representative replaced the touch screen and performed functional tests.No further issues were reported.The investigation is ongoing.A follow-up report will be sent when the investigation is complete.
 
Event Description
Sorin group (b)(4) received a report that the touch screen of an scpc centrifugal pump console was defective.There was no report of patient injury.
 
Manufacturer Narrative
(b)(4) manufactures the centrifugal pump console.The event occurred in (b)(6).This medwatch report is being filed on behalf of livanova (b)(4).Inspection by the livanova service representative identified signs of moisture penetration into the touch screen.This was concluded to be the root cause of the reported issue, which was resolved by replacement of the complete display front panel.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by livanova technician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STOCKERT CENTRIFUGAL PUMP CONSOLE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
SORIN GROUP DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key5089589
MDR Text Key26475599
Report Number9611109-2015-00406
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K020571
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Followup
Report Date 03/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-03-00
Device Catalogue NumberN/A
Device Lot NumberN/A
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/18/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-