BOSTON SCIENTIFIC NEUROMODULATION PRECISION SPECTRA®; SPINAL CORD STIMULATOR Back to Search Results
Model Number SC-2316-50
Device Problems Break (1069); Fracture (1260); High impedance (1291); Kinked (1339)
Patient Problem Inadequate Pain Relief (2388)
Event Date 08/06/2015
Event Type  Injury  
Manufacturer Narrative
Device evaluation indicated that the complaint was confirmed.The lead body had a pronounced kink approximately 18 cm from the proximal end, where the lead appeared to have been sutured.Twelve cables were broken/severed at this spot.The fractured site was 1 cm from the clik anchor setscrew mark.This appeared to have been caused by fatigue possibly coupled with postural changes/movements.The completely severed cables were the reason for the high impedances observed.There were no exposed cables.
 
Event Description
A report was received that the patient had lost stimulation on the left leg.It was noted that all of the left side contacts were showing high impedances.The patient underwent a revision procedure wherein a lead was replaced.Device malfunction was suspected.The patient was reportedly doing well post operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key5089616
MDR Text Key26225830
Report Number3006630150-2015-02368
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Report Date 08/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2016
Device Model NumberSC-2316-50
Other Device ID NumberM365SC2316500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age53 YR
-
-