Model Number N/A
Device Problems Fracture (1260); Difficult to Remove (1528)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 09/03/2015
Event Type  Injury  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: ¿surgical instruments and instrument cases are susceptible to damage for a variety of reasons including prolonged use, misuse, rough or improper handling.¿.
 
Event Description
It was reported patient underwent an initial right shoulder procedure on (b)(6) 2015.During the procedure, a prong fractured off the instrument.The surgeon attempted for forty (40) minutes to remove the fractured prong before continuing with the procedure.The prong remains in the patient's glenoid and cannot be removed.Currently there are no plans for a revision to remove the fractured piece.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSO FORKED RETRACTOR/UK
Type of Device
RETRACTOR
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key5090960
MDR Text Key26284291
Report Number0001825034-2015-04037
Device Sequence Number1
Product Code GAD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Report Date 09/03/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number402852
Device Lot Number153960
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/21/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Disability;
Patient Age74 YR
-
-