• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY Back to Search Results
Catalog Number 03P75-01
Device Problems Failure to Power Up (1476); Battery Problem (2885); Charging Problem (2892)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2015
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).Two analyzers were reported on the initial call.However, analyzer sn (b)(4) was filed as becoming hot to the touch per the complaint.Refer to mdr 2245578-2015-00036 for analyzer sn (b)(4) as it was suspected that only one analyzer would be returned.As a result analyzer (b)(4) was also returned and the investigation confirmed the customer's complaint in which both analyzers failed in the same manner.The investigation was completed on 09/10/2015.The failure was due to a defective tantalum capacitor.
 
Event Description
On 06/18/2015, abbott point of care (apoc) was contacted by a customer who reported that analyzer sn (b)(4) became warm but not hot to touch and does not hold a charge.The customer states that it has happened on multiple downloaders.Customer inserted 9v lithium batteries and the analyzer does not activate.The customer states the analyzer was operating with rechargeable and batteries with a red (fused) battery carrier.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that rechargeable batteries and also a red fused battery carrier was being used.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Type of Device
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer (Section G)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
Manufacturer Contact
linda maczuszenko
400 college road
princeton, NJ 08540
6136885949
MDR Report Key5095202
MDR Text Key26741563
Report Number2245578-2015-00050
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001387
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Report Date 06/18/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P75-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2015
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/22/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-