Catalog Number 341398 |
Device Problem
Metal Shedding Debris (1804)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/28/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device sample has been returned however the investigation report has not been submitted at the time of this report.The manufacturer will continue to monitor and trend related events.
|
|
Event Description
|
Alleged event: metal shavings from the retractor were found inside the patient.The patient's condition was reported as fine.
|
|
Manufacturer Narrative
|
(b)(4).Sap was reviewed for notifications pertaining to incoming inspection, stock checks, and product returns.No concerns were noted with reference to (b)(4).(1) sample of (b)(6) was received for evaluation.Visual examination did not detect any metal shavings.The device was disassembled to find damage to the pinion.Galling was observed on the internal surfaces.The galling is the result of initial damage caused during assembly.The rack may have impacted/damaged the sliding surfaces of the pinion during assembly.Alternately- debris may have been captured in the joint prior to use.Either would result in the galling of the joint.The customer has had this happen twice now; this is a site specific failure.The device still functions as intended when tested under heavy loading.No metal shavings were observed during testing.It is also possible that the metal shavings were introduced during sterilization and may have originated from another device in the load.No functional issues found.
|
|
Event Description
|
Alleged event: metal shavings from the retractor were found inside the patient.The patient's condition was reported as fine.
|
|
Search Alerts/Recalls
|