• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES USA 2.0MM IMF SCREW SELF-DRILLING 12MM; SCREW FIXATION INTRAOSSEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES USA 2.0MM IMF SCREW SELF-DRILLING 12MM; SCREW FIXATION INTRAOSSEOUS Back to Search Results
Catalog Number 201.932
Device Problems Break (1069); Device Or Device Fragments Location Unknown (2590)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/02/2015
Event Type  malfunction  
Manufacturer Narrative
Patient initials are (b)(6).Device broke intra-operatively and was not fully implanted or explanted during the procedure on (b)(6) 2015.Investigation could not be completed and no conclusion could be drawn as no device was returned.Without a lot number, the device history record review could not be requested.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that three (3) intra-medullary fixation screws (12mm) broke during a cerclage procedure on (b)(6) 2015.The first screw, meant for the top of the jaw, broke during insertion.Half of the screw remains in the patient's bone.Two additional screws broke during the procedure.Fragments from one of those screws remain in the patient's bone as well.The heads of the broken screws could not be found and were likely captured by the suction hose.The procedure was completed successfully with no reported time delay.The patient is said to be in good post-operative condition.This report is 3 of 3 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM IMF SCREW SELF-DRILLING 12MM
Type of Device
SCREW FIXATION INTRAOSSEOUS
Manufacturer (Section D)
SYNTHES USA
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES USA
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key5098383
MDR Text Key26628015
Report Number2520274-2015-16092
Device Sequence Number1
Product Code DZL
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
PK010527
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Report Date 09/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number201.932
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-