LIEBEL FLARSHEIM HUT EXT DR FINAL ASSY-STANDARD Back to Search Results
Model Number HUT EXT DR FINAL ASSY-STANDARD
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/10/2015
Event Type  malfunction  
Manufacturer Narrative
Technical service troubleshot with customer their reported a table motion problem and determined that one of the hydraulic pumps had failed.Customer ordered replacement p/n 402066 and after speaking with tech support were able to repair the system themselves, and returned the unit to service.
 
Event Description
Customer reports all table motions failed during an unknown patient procedure.Staff was able to complete the procedure without incident.No reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUT EXT DR FINAL ASSY-STANDARD
Type of Device
HUT EXT DR FINAL ASSY-STANDARD
Manufacturer (Section D)
LIEBEL FLARSHEIM
2111 east galbraith road
cincinnati OH 45237
Manufacturer Contact
david benson
2111 east galbraith road
cincinnati, OH 45237
5139485719
MDR Report Key5100444
MDR Text Key26854513
Report Number1518293-2015-00096
Device Sequence Number1
Product Code IXR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Radiologic Technologist
Report Date 09/10/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHUT EXT DR FINAL ASSY-STANDARD
Device Catalogue Number404008
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse