Model Number 304-20 |
Device Problems
Degraded (1153); High impedance (1291); Detachment of Device or Device Component (2907)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/28/2015 |
Event Type
malfunction
|
Manufacturer Narrative
|
|
|
Event Description
|
During a patient's generator replacement surgery it was noted that the silicone lead material was broken down, looked clear and "slimy," and was observed in the pocket and along the lead.The material had reportedly degraded and was detached from the lead wire.It was noted that the generator had the same material on it.The surgeon inspected the degradation upon explant of the generator and felt there may have been a bend or crease in the lead but was not certain.Pre-surgical diagnostics were noted to be normal with a lead impedance of 1427 ohms.This vns therapy system was completely explanted.The patient's re-implant is pending culture results on the explanted material, which have not been reported to the manufacturer.The explanted devices have been returned to the manufacturer and are currently undergoing product analysis.Attempts for additional relevant information have not been successful to date.
|
|
Event Description
|
The physician reported that the cultures came back negative for infection.Initial visual analysis by the manufacturer revealed abraded openings on the outer and inner silicone tubes.The outer insulation appeared to have been folded outward in one section.The abraded openings found on the outer and inner silicone tubes most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids found inside the outer and inner silicone tubes.Isolated outer insulation abrasions were noted elsewhere on the lead.With the exception of the observed abraded openings, the condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies, beyond the outer and inner tubing openings, were noted.Note that since the electrode array section was not returned for analysis, an evaluation and resulting commentary cannot be made on that portion of the lead.The condition of the outer insulation and accumulation of body fluid inside the lead is the likely cause for the reported appearance of the lead at the time of surgery.There were no performance or any other type of adverse conditions found with the pulse generator.The battery measured 2.822 volts and showed an ifi=no condition.(b)(4).
|
|
Event Description
|
The patient's mothers reported that she suspects the cause for the fluid around the lead was due to the patient making rapid repeated head movements that may have caused a blister around the leads.No additional relevant information has been received to date.
|
|
Search Alerts/Recalls
|