(b)(4).The complaint airvo humidifier was returned to our fisher & paykel healthcare regional office in (b)(4) for evaluation.The device was performance tested and passed all functional tests.The hospital had stated that the airvo flow rate dropped without them adjusting the flow themselves.Based on the description of events it is most likely that the flow had become restricted due to a blockage in the breathing system.The most likely cause for this is kinking of the prongs.If the flow rate drops below a certain limit the airvo will alarm with a "cannot reach target flow" audio and visual alarm.The user then has the choice to accept the lower flow rate or address the problem in order to restore the correct flow rate.No fault was found with the subject airvo.We have concluded that the reported incident was not caused by a malfunction of the airvo itself but rather by a blockage in the flow delivery system, most likely caused by kinking of the prongs.The airvo 2 user manual provides the folllowing warnings: - do not block the flow of air through the unit and breathing tube.The unit cannot reach the target flow setting: - check the heated breathing tube or patient interface for blockage.The manual also states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases." and that "the unit is not intended for life support.".
|
A hospital in (b)(6) reported that a child with pneumonia was connected to an airvo 2 humidifier, with a flow of 12 l/min and 45% o2 after four days of therapy, it was observed that the child's saturation values had decreased to less than 80%.It was found that the airvo was only showing a flow of 2 l/min on the display, even though the set flow was still 12 l/min.The child was removed from the airvo and put on standard o2 therapy.The child then recovered quickly.
|