• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP Back to Search Results
Model Number 16402
Device Problem Component Missing (2306)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/03/2015
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The field service representative (fsr) found a note attached on the roller pump that read: "missing tongue".
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the pump tongue was missing from the roller pump.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint was confirmed.The customer was unable to provide further information as to why the pump tongue is missing, such as damage or removal to install on another pump.The missing tongue was replaced in in-house service.Released testing was not performed at this time due to additional parts not available to complete repair.The product will be brought to manufacturers specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE - SARNS ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key5104009
MDR Text Key26956505
Report Number1828100-2015-00832
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Followup
Report Date 10/21/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/25/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received10/21/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-