During device implantation, initial implant impedance measurements were acceptable.During the final placement of the neuroregulator, a low impedance alarm was indicated on the posterior lead.The low impedance alarm, in accordance with intended product design, prevented initiation of therapy.Because the procedure had been completed to the point where medical adhesive covered the lead connection on the neuroregulator, both the neuroregulator and posterior lead were removed and replaced with alternate components in accordance with standard practice.Implantation of alternate components resolved the low impedance alarms.The implantation procedure completed successfully without adverse impact to the patient.Final impedance measurements were somewhat low, so the impedance amplitude setting was not lowered as typically done at implant, to ensure better impedance measurement accuracy.
|