• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION PROPAQ MD DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number PROPAQ MD
Device Problems Loss of Power (1475); Self-Activation or Keying (1557); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while monitoring ecg, spo2 and nibp on a patient (age & gender unknown), the device prompted a "system reset" message and switched on and off by itself three times.Complainant indicated that the device powered back on and they continued to treat the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.Please reference medwatch report 1220908-2015-02506 for a similar incident with the same device.
 
Manufacturer Narrative
Zoll medical corporation evaluated the device and the reported malfunction was not replicated or confirmed.The device was put through extensive testing without duplicating the malfunction.Review of the device activity logs provided evidence of the customer report.The device's multi-function cable and defib cable receptacle were replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPAQ MD DEFIBRILLATOR
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key5107851
MDR Text Key27004935
Report Number1220908-2015-02471
Device Sequence Number1
Product Code DRT
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K100654
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation Other
Type of Report Followup
Report Date 09/09/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberPROPAQ MD
Device Catalogue NumberPROPAQ MD
Other Device ID Number00847946017705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/07/2015
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/28/2015
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received11/18/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-