• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - MALTA ACCESS; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - MALTA ACCESS; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number FMC9673
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection noted that the tubing of the set had been crushed by the pouch sealing machine.This is due to a manufacturing issue.Sealing machines have since had sensors to detect tubing that is sticking out of the packaging in order to prevent this occurrence.A capa has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a blood administration pump set had a "manufacturing defect on the tube".This was observed before use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - MALTA
a47 industrial estate, malta b
marsa
MT 
Manufacturer (Section G)
BAXTER HEALTHCARE - MALTA
a47 industrial estate, malta b
marsa
MT  
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key5108678
MDR Text Key26940307
Report Number1416980-2015-37426
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeSN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Report Date 09/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberFMC9673
Device Lot Number14K06V615
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/28/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-